An economical solution to your diagnostic blood work needs in smaller clinics and cost-sensitive healthcare environments. Low initial cost, minimal training requirements, self-prompting interactive operation. Advanced electronics and software available. Obtain fast results from small samples. Large test menu for over 50 assays, internal printer with graphics capability, Integral heating block, RS232 interface. |
|
| Linear Measurement Range |
-0.20 to 2.50 Absorbance Units (A) |
| Photometric Accuracy |
± 1% of the reading + 0.005 A |
| Stability |
drift of no more than 0.005 A in 8 hours/bichromatic |
| Light Source |
tungsten halogen lamp, with lamp saver feature |
| Wavelengths |
340, 405, 450, 505, 545, 600 nm (alternate filters from 340 to 750 nm can be custom ordered) |
| Type of Filter |
multiple cavity hard coat interference, 10nm half bandpass |
| Tube Size |
12 mm round, borosilicate glass tube |
| Minimum Fill Volume |
1 ml for 12 mm tube |
| Display |
alpha numeric, 16 character LCD |
| Printer |
thermal, dot matrix, 20 characters per line, plus graphics |
| Keyboard |
16-key, domed membrane switch, enunciating |
| Power Requirements |
115V or 230V AC, 50-60 Hz (switch selectable) |
| Microprocessor |
Z80A or Z180 |
| Speed |
reads, calculates and prints results in 3 seconds/tube |
| Calculation Modes |
single point calibration by standard factor, uptake, multipoint calibration with point to point curve fit, rate by standard or factor (batch or singularly) |
| Test Menu |
More than 50 open channels to store tests. Stores all parameters including wavelengths, calculations, unit codes, linear and normal ranges, rate timing, standard values, test names, and previous standard curve |
| Temperature |
37°C, block stays on, cell has on/off switch |
| Enclosure |
painted flame retardant ABS plastic cover with metal base |
| Dimensions |
approx. 9x12x3 in. (23x30x8 cm), 15 lbs. (6.8 kg) |
| Software (Optional) |
Stat Tracks is a data management software that connects your BMD 1940 Plus analyzer to your PC to allow simple yet powerful tracking and reporting capabilities. With Stat Tracks™, you can store patient lists, setup and run worklists, see curves, print reports, and collect QC data. Stat Tracks, employs a graphical user interface which makes it very simple to use. All functions are menu-driven. |
| Certifications |
Health Canada TPP Lic. #28622 Device Class 2; CE Mark, and FDA approved |